Abstract

The revelation that intravenous sodium fluorescein is not all that it might seem to be may be a significant finding in the light of the adverse reactions to fluorescein that have been previously reported. Analysis of commercially prepared intravenous sodium fluorescein by mass spectroscopy and nuclear magnetic resonance has indicated that an industrial solvent used in the manufacturing process has not been eliminated. Dimethyl formamide is an industrial solvent with a maximum acceptable exposure level of 10 parts per million for dermal contact. It has been found in quantities of 5000 parts per million in the fluorescein for intravenous use. This investigation was prompted by a significant increase in the adverse reactions in patients receiving intravenous fluorescein in the retinal photographic unit at the Manchester Royal Eye Hospital.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.