Abstract

A pneumatic sampling switch which can be adapted to the outlet of a thermogravimetric analyzer furnace with a horizontal purge gas flow was developed. This system allows a repetitive introduction of short sampling pulses of thermogravimetric analyzer evolved gases to a quadrupole mass spectrometric detector. System performance was evaluated through measuring the level of repeatability under different operation conditions. In addition, pulsed sampling thermogravimetry/mass spectrometry introduced no measurable quantitative or qualitative errors. This system can be used for monitoring thermooxidative decomposition reactions, without accelerating the deterioration of quadrupole mass spectrometric detector performance. Furthermore, it can be used as an inlet system for a future on-line TG/GC/MS system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.