Abstract
Device interrogation of an asymptomatic pacemaker recipient (St Jude Medical Accent DR) following ablation of atrial fibrillation (AF), revealed 29 auto mode switch (AMS) episodes, thus raising the suspicion of asymptomatic AF recurrences. However, all stored AMS episodes were due to repetitive non-reentrant ventriculoatrial synchrony (RNRVAS). In St Jude Medical devices, …
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.