Abstract
PurposeTo assess the repeatability of intraocular pressure (IOP) measured with the Corvis ST (CST) and the Ocular Response Analyzer (ORA).MethodsA total of 141 eyes from 141 subjects were studied, including 35 healthy eyes and 106 glaucomatous eyes. All subjects underwent IOP evaluations with Goldmann applanation tonometer, CST, and ORA. With CST, biomechanical corrected IOP (bIOP) was calculated; bIOP is purported to be less dependent on biomechanical properties. For ORA, corneal-compensated intraocular pressure (IOPcc) and Goldmann-correlated IOP (IOPg) were derived. The repeatability of the various IOP values was assessed using the coefficient of variance (CV) and the intraclass correlation coefficient (ICC).ResultsThe CV with bIOP (5.5 ± 3.1: mean ± standard deviation) was significantly smaller than the CVs measured with IOPg (7.3 ± 4.3) and IOPcc (7.2 ± 4.4). ICC values were 0.90, 0.80, and 0.86 with IOPg, IOPcc, and bIOP, respectively.ConclusionsThe bIOP showed a better prevision and repeatability for IOP measurement.Translational RelevanceThe bIOP measurement from CST had a better reproducible than IOPcc measurement from ORA.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.