Abstract

To evaluate the repeatability of optic nerve head (ONH) measurements by spectral-domain optical coherence tomography. Three scans were acquired in 32 healthy subjects during one session. Using Cirrus HD-OCT (Carl Zeiss Meditec, Dublin, CA), the cup-to-disc ratio (CDR), disc area, rim area, cup volume, and horizontal and vertical CDRs were investigated. Repeatability was assessed by the coefficient of variation (COV), the test-retest intrasession variability, and the intraclass correlation coefficient (ICC). Good repeatability was achieved for all parameters, with a COV of 4.23% or less and ICCs of 0.98 or greater for all measurements. Test-retest intrasession variability was 0.024 for CDR, 0.121 mm(2) for disc area, 0.087 mm(2) for rim area, 0.017 mm(3) for cup volume, and 0.048 for horizontal and vertical CDR. In healthy eyes, Cirrus HD-OCT provides repeatable measurements of ONH parameters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.