Zebrafish (Danio rerio) are widely used as an experimental model for a wide range of retinal diseases. Previously, optical coherence tomography (OCT) was introduced for quantitative analysis of the zebrafish cone photoreceptor cell mosaic; however no data exists on the intersession reproducibility or intrasession repeatability of such measurements. We imaged 14 wild-type (WT) fish three times each, with 48h between each time point. En face images of the UV cone mosaic were generated from the OCT volume scans at each time point. These images were then aligned and the overlapping area cropped for analysis. Using a semiautomated cone-counting algorithm, a single observer identified each cone to calculate the cone density for every image, counting each image twice (84 total counts). The OCT cone density measurements were found to have an intersession reproducibility of 0.9988 (95% CI=0.9978-0.9999) and an intrasession repeatability of 136.0±10.5 cones/mm2 (about 0.7%). Factors affecting image quality include gill movement during acquisition of the OCT volume and variable inclusion of non-UV cone mosaics in the contours used to generate the en face images.