Abstract

We evaluate the inter- and intraobserver variability of a telemanipulated femur fracture reduction system using a joystick device. Five examiners performed virtual reduction of 3D femur fracture models on two separate occasions. We assessed the inter- and intraobserver variability for the final alignment and reduction. The average difference between testing rounds was only 0.3mm for overall displacement and 0.5° for overall rotation. There was an average time reduction between rounds of 11.7s. The mean differences in overall displacement between examiners ranged between 0.2 and 0.9mm; between 0.2° and 3.2° for overall rotation; and between 9 and 82s for time to reduction. The time required to complete the reduction did not have a significant effect on the overall displacement or rotation of the final model. Telemanipulated fracture reduction is a reliable and reproducible technique, which does not require extensive training.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.