Abstract

To evaluate the repeatability of corneal biomechanical parameters in normal and keratoconus eyes, and explore factors that affects the repeatability, and further assess the diagnostic ability of new parameters. Seventy-seven keratoconus eyes of 47 patients and 77 right eyes of 77 normal subjects were recruited in current study. All participants received three repeated measurements with 2 to 5 minutes interval. The interclass correlation coefficient (ICC), Cronbach’ α and repeatability coefficient (RC) were evaluated. The liner regression analysis was used to identify factors that affect the repeatability, and linear mixed effects model was performed to compare the parameters differences. The receiver operating characteristic (ROC) curve was used to evaluate the diagnostic ability of new parameters. Eighteen parameters in normal eyes and twenty-two parameters in keratoconus eyes showed excellent repeatability (ICC ≥ 0.90). Age, axial measurement (AL), spherical equivalent, astigmatism, gender, mean keratometry (Kmean), intraocular pressure (IOP) and central corneal thickness (CCT) could affect the repeatability of new Corvis ST parameters. Compared with normal eyes, the Ambrósio’s Relational Thickness horizontal (ARTh), biomechanical corrected IOP (bIOP), stiffness parameter at first applanation (SP A1) were low and the Max Inverse Radius, deformation amplitude (DA) Ratio Max [2 mm], Pachy Slope, DA Ratio Max [1 mm], Integrated Radius and Corvis Biomechanical Index (CBI) were high in keratoconus eyes (All P < 0.05). Both ARTh and CBI had high Youden index (0.870), and the corresponding cut-off values were 379.29 and 0.44. The repeatability of Corvis ST parameters was acceptable both in normal and keratoconus eyes, and new parameters could effectively diagnose keratoconus eyes from normal eyes.

Highlights

  • To evaluate the repeatability of corneal biomechanical parameters in normal and keratoconus eyes, and explore factors that affects the repeatability, and further assess the diagnostic ability of new parameters

  • The corneal biomechanical properties were important in evaluating the ophthalmic examinations before refractive surgery and detecting keratoconus

  • The current study found that the repeatability of 39 Corvis ST parameters in normal eyes and keratoconus eyes were acceptable and new Corvis ST measurements could help to diagnose keratoconus eyes in clinical application

Read more

Summary

Introduction

To evaluate the repeatability of corneal biomechanical parameters in normal and keratoconus eyes, and explore factors that affects the repeatability, and further assess the diagnostic ability of new parameters. The Ambrósio’s Relational Thickness horizontal (ARTh), biomechanical corrected IOP (bIOP), stiffness parameter at first applanation (SP A1) were low and the Max Inverse Radius, deformation amplitude (DA) Ratio Max [2 mm], Pachy Slope, DA Ratio Max [1 mm], Integrated Radius and Corvis Biomechanical Index (CBI) were high in keratoconus eyes (All P < 0.05). Both ARTh and CBI had high Youden index (0.870), and the corresponding cut-off values were 379.29 and 0.44. We investigated the differences of parameters between normal and keratoconus eyes, and further evaluated the diagnostic ability of new parameters in detecting keratoconus from normal eyes

Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.