Abstract

BackgroundPulsed-field ablation (PFA) is a nonthermal energy source for ablation of cardiac arrhythmias. This study investigated the prospective outcomes of a novel PFA generator in conjunction with a commercially available, contact force-sensing, focal ablation catheter. ObjectiveThe purpose of this study was to assess the feasibility, safety, and lesion characteristics of point-by-point PFA in consecutive patients undergoing repeat ablation of atrial fibrillation (AF). MethodsThe study involved reisolation of pulmonary veins (PVs) with electrical reconnection and the creation of an anterior line (AL) in patients with anterior substrate or durable pulmonary vein isolation (PVI). ResultsIn 24 patients (46% female; mean age 67 ± 10 years; 67% persistent AF), successful reisolation of 27 of 27 reconnected PVs (100%) was performed. In 19 patients, AL ablation was performed, with bidirectional block in 16 (84%), median ablation time 26 [21, 33] minutes, and first-pass bidirectional block in 13 patients (68%). Acute AL reconduction occurred in 8 of 19 patients (42%). Among these 8 patients, a subsequent sustained block of the AL was achieved in 5 (63%). Ultra-high-density electroanatomic mapping revealed homogeneous but relatively large low-voltage areas in the ablated regions. Median procedural, left atrial dwell, and fluoroscopy times were 100 [90, 109] minutes, 83 [75, 98] minutes, and 10 [8, 13] minutes, respectively. No major or minor complications occurred. ConclusionThis study demonstrated feasibility, acute efficacy, and safety of point-by-point PFA for repeat PVI and AL ablation. Further studies are warranted to assess the long-term durability and comparison with established ablation methods.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call