Abstract

The diagnosis of the different forms of thalassemia is one of the important applications of analysis of globin chains. These analyses are done by high performance liquid chromatography (HPLC) using a MONO-S cation exchange column and ether is used for washing the globin powder in the final step. The presence of peroxide impurities in ether could change the structure of the globin chains. In the chromatograms, these modified globins appear as separated peaks next to the unmodified globin peaks. In these cases, the alpha/beta ratio exceed by artifact the correct value. Our study demonstrates that diagnostic centers should ensure that the ether they use is pure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.