Abstract

A simple, fast and sensitive spectrophotometric method is developed for removal, preconcentration and determination of trace amounts of picric acid in water samples. Magnetic iron oxide nanoparticles (MIONPs) were synthesized and characterized by transmission electron microscopy (TEM). The magnetic nanoparticles were coated with cetyltrimethylammonium bromide (CTAB) and were applied for fast separation, preconcentration and spectrophotometric determination of picrate anion (the ion of picric acid) in an aqueous solution. The separation, preconcentration procedure is fast and will be completed in 2min. Methanol is used for desorption of adsorbed picrate anion. The effects of important parameters such as pH of aqueous medium, CTAB dosage, adsorbent amount, temperature, electrolyte concentration, desorbing solvent and interfering ions on the adsorption of picrate anion were investigated. The method showed good linearity for the determination of picric acid in the concentration range of 0.02–1.00μgmL−1 with a regression coefficient of 0.999. The limit of detection (LOD) is obtained to be 0.007μgmL−1. The relative standard deviation (RSD) for 0.03μgmL−1 and 0.8μgmL−1 of picric acid were 3.98% and 1.97% respectively. Picric acid was separated, preconcentrated and determined successfully in spiked samples of Karoon River water.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call