The development of solid nails has not eliminated the risk of occasional nail failure. It has been suggested that subsequent intra-medullary removal of solid nail fragment may be virtually impossible and thus major surgery is required for their revision. We report two cases of hardware failure of a solid intra-medullary nail (Synthes solid femoral and tibial nail) and describe their successful intra-medullary removal using the Synthes extraction kit.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.