Abstract

Purpose: Maldeployment of the Supera stent system can result in primary technical failure, inferior primary patency, and poorer patient outcomes. The purpose of this article is to present a case series illustrating the conditions required to perform optimal stent deployment, and if necessary, undertake successful stent removal following maldeployment. Technique: Two key failures of effective Supera deployment are elongation and invagination. Several technical factors should be considered to reduce the risk of maldeployment: aggressive target vessel predilation, the use of multiple fluoroscopic views, slow deployment with controlled forward pressure applied on the delivery device, “sandwich packing” of the stent above and below target lesions, and the “pulling back” of invagination. To successfully retrieve a partially deployed stent, 3 factors should be considered: the percentage of the stent already deployed, the distance from the distal tip of the introducing sheath to the proximal extent of the deployed stent, and the severity of proximal vessel disease. The higher these factors, the greater the risk of stent detachment and failed retrieval. Conclusion: In this series of 6 cases of maldeployment, the removal of a partially deployed Supera stent appeared to be feasible and safe, with success dependent on selected technical and anatomical considerations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.