Abstract

Background It has been reported that intravoxel incoherent motion (IVIM) diffusion magnetic resonance imaging (MRI) scan-rescan reproducibility is unsatisfactory. Purpose To study IVIM MRI parameter reproducibility for liver parenchyma after the removal of motion-contaminated and/or poorly fitted image data. Material and Methods Eighteen healthy volunteers had liver scans twice in the same session to assess scan-rescan repeatability, and again in another session after an average interval of 13 days to assess reproducibility. Diffusion-weighted images were acquired with a 3-T scanner using respiratory-triggered echo-planar sequence and 16 b-values (0-800 s/mm2). Measurement was performed on the right liver with segment-unconstrained least square fitting. Image series with evidential anatomical mismatch, apparent artifacts, and poorly fitted signal intensity vs. b-value curve were excluded. A minimum of three slices was deemed necessary for IVIM parameter estimation. Results With a total 54 examinations, six did not satisfy inclusion criteria, leading to a success rate of 89%, and 14 volunteers were finally included for the repeatability/reproducibility study. A total of 3-10 slices per examination (mean = 5.3 slices, median = 5 slices) were utilized for analysis. Using threshold b-value = 80 s/mm2, the coefficient of variation and within-subject coefficient of variation for repeatability were 2.86% and 3.36% for Dslow, 3.81% and 4.24% for perfusion fraction (PF), 18.16% and 24.88% for Dfast; and those for reproducibility were 2.48% and 3.24% for Dslow, 4.91% and 5.38% for PF, and 21.18% and 30.89% for Dfast. Conclusion Removal of motion-contaminated and/or poorly fitted image data improves IVIM parameter reproducibility.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call