Abstract

Stent dislogement or migration is not a rare complication. Its incidence varies from 1.4% to 8% of the cases. When stent migration occurs, the usual treatment or management is to implant stent in the distal peripheral artery or proximal coronary artery. But, probably the best treatment is safe retrieval of the dislodged stent. During the transfemoral coronary intervention, undeployed stent can be more easily retrieved into the guiding catheter with relatively larger guiding catheter luminal space. Also, larger sheath can be changed more easily when the retrieval of stent is difficult. In the transradial coronary stenting, usual size of guiding catheter is 6 Fr, makes it difficult to retrieve relatively bulky stent when deployment of stent fails. We report 2 cases of stent dislodgment during transradial coronary stenting which can be successfully removed by using myocardial biopsy forcep through 8 Fr sheath. These combination could be one of the valuable methods, especially during transradial stenting. (Korean Circulation J 2000;30(10):1312-1315)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.