Abstract

Remote Vision-Based digital Patient Monitoring (VBPM) of pulse (PR) and respiratory rate (RR) was set up in six single rooms in an acute medical and an orthopaedic ward. We compared 102 PR and 154 RR VBPM measurements (from 27 patients) with paired routine nurse measurements. VBPM measurements of RR were validated by reviewing video footage. Nurse measurements of RR were often 16-18 breaths/minute, and did not match VBPM RR (overestimating at low RR and underestimating at high RR). Nurse measurements of pulse were on average 3.9 beats per minute greater than matched VBPM measurements. VBPM was unobtrusive and well accepted.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.