Abstract

This experimental investigation was aimed at verifying the feasibility and safety of a remote-controlled wireless chest drainage system for in-hospital, outpatient, or domiciliary management of prolonged air leaks. A commercially available digitalized portable vacuum unit was adapted to incorporate a unit for telemetric data transmission and controlling suction levels through Global System for Mobile Communication (GSM) technology. The system was tested in 1 patient after right upper lobectomy and wedge resection of the right lower lobe. The air leaks were monitored in the High Dependency Unit, on the thoracic floor, and after admission to an outside guest house. The system showed dependable monitoring and alteration of suction levels equally in the 3 environments and at different stages of the postoperative course. A remote-controlled wireless chest drainage system is technically feasible and appears dependable. Its use in the management of large series of patients will contribute to the refinement of indications and technical troubleshooting.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.