Abstract
High-throughput processing of parallel-beam X-ray tomography at synchrotron facilities is lacking a reliable and robust method to determine the center of rotation in an automated fashion, i.e. without the need for a human scorer. Well-known techniques based on center of mass calculation, image registration, or reconstruction evaluation work well under favourable conditions but they fail in cases where samples are larger than field of view, when the projections show low signal-to-noise, or when optical defects dominate the contrast. Here we propose an alternative technique which is based on the Fourier analysis of the sinogram. Our technique shows excellent performance particularly on challenging data.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.