Abstract

Our purpose in this work was to assess the reliability of the calibration coefficient for magnetic resonance water proton chemical shift temperature mapping. Over a six month period, the calibration coefficient was measured 15 times in several different phantoms. A highly linear relationship between water proton chemical shift and temperature change was found. The average temperature calibration coefficient determined from all studies was 0.009+/-0.001 ppm/degrees C. Four of the 15 studies were conducted on the same day using the same phantom. The average temperature calibration coefficient of these four studies was 0.0096+/-0.0001 ppm/degrees C.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.