Abstract

Early and accurate diagnosis is crucial for preventing the spread of SARS-CoV-2 infection. The rapid antigen test was developed for testing infection, and it was necessary to assess its performance before widespread use in Tunisia. To evaluate the effectiveness of a rapid antigen test for the detection of SARS-CoV-2 in nasopharyngeal swabs in Tunisia. Nasopharyngeal samples were taken from COVID-19 suspected cases between October and December 2020 and tested using the Standard Q COVID-19 Ag test (SD-Biosensor, Republic of Korea) and real-time reverse transcription polymerase chain reaction (RT‑PCR). Overall, 4539 patients were tested. Of the total study population (N = 4539), 82.5% of positive samples remained positive with the rapid antigen test, while 20.2% (470/2321) of samples that were negative with rapid antigen test were confirmed positive with RT-PCR, giving a negative predictive value of 79.8% for the rapid antigen test. The sensitivity and negative predictive value of the rapid antigen test were 70.2% and 65.8%, respectively. These results improved to 96.4% and 92.8%, respectively, when considering the cycle threshold value by RT-PCR below 25. Although the rapid antigen test was less sensitive than RT-PCR, its ability to rapidly detect individuals with high viral loads makes it suitable for use during an epidemic.