Abstract

To evaluate the reliability of the Osstell implant stability quotient (ISQ) and Penguin resonance frequency analysis (RFA) devices in measuring implant stability. Forty implants were embedded in self-curing acrylic resin, soft-lining material, polyvinyl siloxane impression material, and polycarboxylate cement (n = 10). After the initial setting times were completed, the stability of each implant was measured with Osstell ISQ and Penguin RFA with 3 repeated measurements. The "intraclass correlation coefficient" evaluated the correspondence between the measurements (P < 0.05). Polyvinyl siloxane impression material had lower ISQ values than soft-lining material, self-curing acrylic resin, and polycarboxylate cement in both devices (P < 0.05). The intraclass correlation was 1.00 in self-curing acrylic resin and 0.48 in polycarboxylate cement (P < 0.05) for Osstell. This value was 0.95 in self-curing acrylic resin and 0.38 in polycarboxylate cement (P < 0.05) for Penguin. There was no correlation between the repeated measurements in soft-lining material and polyvinyl siloxane impression material for both devices (P > 0.05). The repeatability was 0.90 for Osstell and 0.60 for Penguin (P < 0.05). Osstell ISQ and Penguin RFA are reliable only when the implants are embedded in stiff materials. Osstell ISQ is more reliable than Penguin RFA.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.