Abstract

In order to investigate the reliability of the Inductively Coupled Plasma Atomic Emission Spectrometer (ICP-AES) for trace element analysis of biological materials, we have carried out extensive investigations on human plasma, using an Applied Research Laboratory's ICP-AES. When we aspirated the untreated plasma into the spectrometer, we obtained unreliable and nonreproducible results. However, when we pretreated the plasma by wet digestion (to destroy all the organic material), we achieved reproducible and consistent results. It is, therefore, suggested that biological samples should be pretreated, preferably by wet digestion, before being aspirated into the ICP-AES for analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.