Abstract
The reliability of reusing the sensitized thermoluminescence (TL) for LiF (TLD-100) dosimeters have been investigated. It has been found that reusing the same sensitized LiF dosimeters may influence the sensitivity of the dosimeters even under the same experimental conditions when the temperature readout range does not include the high-temperature TL signal at T > 505K. But, if the high-temperature TL signal is included in the temperature readout range, it has been found that the ultra-high temperature TL signal significantly increases with increasing the number of reusing the dosimeters. After almost 7 times of reusing the sensitized dosimeters, the total area under the glow-curve becomes stable and independent of the number of reusing the dosimeters. It has been also found that the TL signal located between the low-temperature TL signal and the ultra-hightemperature TL signal was almost stable over all the number of reusing the dosimeters. For all the lowtemperature peaks, the kinetics parameters of the sensitized LiF dosimeters have not to be influenced by reusing the same dosimeters. However, the activation energy of peak 5 which lies at temperature range (460-500K) has been influenced by the readout temperature range and reusing the dosimeters. The typical value for this glowpeak has been obtained in the case which the high-temperature TL signal has been included in the temperature readout.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.