Abstract

Purpose: The aim of this study is to determine intra and inter-session repeatability and reliability of OHM 3000 for reporting static and dynamic pressures in healthy Indian population. Methods: Using OHM 3000, twenty-eight healthy adults who did not present any gait abnormalities or musculoskeletal conditions were assessed one week apart for static and dynamic pressure. Bipedal stance test was used for assessing static variables- maximum pressure, average pressure, contact area, and weight distribution; whereas dynamic test was used for assessing dynamic variables- maximum pressure, average pressure, and contact area. Each participant performed three trials under supervision of each investigator. Results: Reliability was devised using ICC values and coefficient of variance for static and dynamic variables. ICC values for static and dynamic variables were reported to range between 0.88 to 0.94 and 0.61 to 0.98 respectively. Covariance values were reported to be lower than 13% in static and dynamic pressures recorded by both investigators. Conclusion: The study reports that OHM 3000 is a reliable device for assessing static and dynamic pressure in healthy adults demonstrating moderate to good reliability. KEYWORDS: Gait, Foot, Plantar Pressure.