Abstract

Dual-energy X-ray absorptiometry (DXA) of the lateral distal femur (LDF) has been suggested for patients with metal implants or joint contractures preventing DXA scanning at conventional anatomical sites. This study assessed variability in LDF DXA measures due to repeat scanning using data from 5 healthy young adults who had 3 unilateral scans with repositioning between scans. Variability due to image analysis was evaluated in 10 children who underwent bilateral LDF scans with each scan being analyzed 3 times by 2 raters. Regions of interest (ROIs) were defined in the anterior distal metaphysis (R1), metadiaphysis (R2), and diaphysis (R3) as described previously. An additional region (R4) was defined in the metaphysis similar to R1 but centered in the medullary canal. Variability was consistently lower for bone mineral density than for bone mineral content and bone area; R4 was more repeatable than R1; and variability because of repeat scanning was negligible. These results suggest that DXA measures of the LDF are reliable and may be useful when standard DXA measures cannot be obtained, but it is recommended that a central, rather than anterior, ROI be used in the metaphysis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.