Abstract

To determine the reliability of Tono-Pen XL in measuring intraocular pressure (IOP) over the amniotic membrane patch in human eyes. Ten healthy volunteers (nine males and one female) participated in this study. After topical anesthesia, IOP was measured using the Tono-Pen XL before and after the application of a single-layer amniotic membrane patch in the right eye and double-layer amniotic membrane patch in the left eye. We found no statistical difference between the IOP measured with and without a single layer amniotic membrane (P = 0.808). Nor did we find any statistical difference between the IOP measured with and without double-layer amniotic membrane patch (P = 0.813). The average thickness of the single- and the double-layer amniotic membrane patch measured by pachymetry was 131.9 +/- 31.4 microm (mean +/- SD) and was 246.1 +/- 69.3 microm (mean +/- SD), respectively. This study showed that it is reliable to measure IOP by Tono-Pen XL over single or a double layer of amniotic membrane patch on healthy human eyes. Further studies are required to determine the accuracy of IOP measurement over amniotic membrane patch on diseased corneas.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.