Abstract

The reliability of glucose test strips and reflectance meters after storage in tropical conditions was assessed. In a first experiment, sealed bottles containing glucose test strips were stored for 1 month in closed sterilising ovens at 40°C, one in dry (0% humidity) and the other in wet (75% humidity) conditions. Results were compared to measurements obtained with glucose test strips stored under standard Western conditions, in non-diabetic and diabetic patients, using the same standard reflectance meter. A second experiment was performed similarly, comparing the values obtained with reflectance meters stored in different conditions, using standard glucose test strips. Reproducibility was assessed on the basis of duplicate measurements for each condition of storage. We observed that temperature and humidity did not significantly affect the glucose test strips (1st experiment), while a slight underestimation of capillary blood glucose was found using reflectance meters stored at 50°C, irrespective of the humidity (2nd experiment). All the correlation coefficients between duplicates were > 0.975 irrespective of the conditions of storage for glucose test strips and reflectance meters. Therefore, glucose test strips and reflectance meters appear fairly reliable at the grass-root level in the tropics, if used carefully.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.