Abstract
We tested the accuracy, sensitivity, and reproducibility of a new lung water computer, based on the thermal conductivity technique, in 22 anesthetized closed-chest ventilated sheep with different treatments: 1) controls (n = 8), 2) 0.05 ml/kg of oleic acid + 100 ml/kg of lactated Ringer solution (n = 6), and 3) airway instillation of saline [3.1 +/- 1.3 (SD) g/kg, n = 8]. After 4 h, we determined the extravascular lung water gravimetrically. We found a significant overall correlation between the final extravascular lung thermal volume and the gravimetric extravascular lung mass (P < 0.001). Although the average ratio of extravascular lung thermal volume to extravascular lung mass was 0.97 +/- 0.25 ml/g for all groups, the computer overestimated extravascular lung mass in controls by 10% (17 g) and underestimated it in sheep with oleic acid by 15% (95 g) and in sheep with airway instillation by 8% (37 g). The computer also underestimated the small quantities of saline placed via the airway in the alveolar space by 75% (61 g). Reproducibility of three consecutive measurements was 4.3% (SE). We conclude that the thermal conductivity technique has an ability to detect the baseline extravascular lung mass but has a poor ability to detect an accurate increment of the extravascular lung water under poor tissue perfusion in anesthetized ventilated sheep.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of applied physiology (Bethesda, Md. : 1985)
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.