Abstract

We tested the accuracy, sensitivity, and reproducibility of a new lung water computer, based on the thermal conductivity technique, in 22 anesthetized closed-chest ventilated sheep with different treatments: 1) controls (n = 8), 2) 0.05 ml/kg of oleic acid + 100 ml/kg of lactated Ringer solution (n = 6), and 3) airway instillation of saline [3.1 +/- 1.3 (SD) g/kg, n = 8]. After 4 h, we determined the extravascular lung water gravimetrically. We found a significant overall correlation between the final extravascular lung thermal volume and the gravimetric extravascular lung mass (P < 0.001). Although the average ratio of extravascular lung thermal volume to extravascular lung mass was 0.97 +/- 0.25 ml/g for all groups, the computer overestimated extravascular lung mass in controls by 10% (17 g) and underestimated it in sheep with oleic acid by 15% (95 g) and in sheep with airway instillation by 8% (37 g). The computer also underestimated the small quantities of saline placed via the airway in the alveolar space by 75% (61 g). Reproducibility of three consecutive measurements was 4.3% (SE). We conclude that the thermal conductivity technique has an ability to detect the baseline extravascular lung mass but has a poor ability to detect an accurate increment of the extravascular lung water under poor tissue perfusion in anesthetized ventilated sheep.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.