AbstractThe material and component requirements for ceramic femoral ball heads in total hip replacement systems are described and briefly discussed. Reliability is correlated to the homogeneity of the material in terms of microstructure and Weibull modulus, and to the design of the components. Stress distribution can be analysed by finite element analysis. Based on controlled and specialised processing and the stress analysis data the design can be optimised. Proof testing, as part of final inspection, is an efficient method for eliminating components having strengths at the low end of the strength distribution spectrum.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.