Abstract

To assess repeatability and reproducibility, to determine normative data, and to investigate the effect of age-related macular disease, compared with normals, on photostress recovery time measured using the Eger Macular Stressometer (EMS). The study population comprised 49 healthy eyes of 49 participants. Four EMS measurements were taken in two sessions separated by 1 h by two practitioners, with reversal of order in the second session. EMS readings were also taken from 17 age-related maculopathy (ARM), and 12 age-related macular degeneration (AMD), affected eyes. EMS readings are repeatable to within +/-7 s. There is a statistically significant difference between controls and ARM affected eyes (t = 2.169, p = 0.045), and AMD affected eyes (t = 2.817, p = 0.016). The EMS is highly specific, and demonstrates sensitivity of 29% for ARM, and 50% for AMD. The EMS may be a useful screening test for ARM, however, direct illumination of the macula of greater intensity and longer duration may yield less variable results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.