Abstract

PurposeTo evaluate the reliability and validity of the femorotibial mechanical axis angle from radiographs in the weight bearing (WB) and supine positions compared with navigation-measured values.Materials and MethodsSixty-eight cases of navigation-assisted total knee arthroplasty (TKA) were included. The pre- and postoperative whole leg radiographs (WLRs) in WB and supine positions were compared with the initial and final navigation values.ResultsThe mean mechanical axis angle from the preoperative WBWLR and navigation were not statistically different (p=0.079) and were correlated strongly with each other (intraclass correlation [ICC], 0.818). However, on postoperative measurements, although the WBWLR and navigation values were not different (p=0.098), they were not correlated with each other (ICC, 0.093). The standard error of measurement was 1.8°±3.6° for the preoperative WBWLR and 2.5°±4.8° for the postoperative WBWLR. The validity that was determined by the Bland-Altman plot was not acceptable for both pre- and postoperative measurements.ConclusionsThe preoperative WBWLR could provide accurate but not precise measurement value of the femorotibial mechanical axis angle for navigation-assisted TKA, and postoperative measurements in navigation were not comparable with radiographic measurements. The lack of agreement was found between the radiographic and navigation measurements of the coronal alignment regardless of pre- or postoperative evaluation although the accuracy was found acceptable.Level of EvidenceLevel 4

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call