Abstract
The reliability and predictive value of the low-pulse-rate electrical auditory brainstem response (EABR) threshold and the high-pulse-rate psychophysical threshold used in electrode mapping of adult cochlear implant patients was assessed. Eight patients, all implanted with the Nucleus multi-channel cochlear implant and experienced users, participated in this study. The relationship between like pulse rates of the EABR and psychophysical thresholds was assessed and showed high correlations of 0.90 or higher. When the relationship between low-pulse-rate EABRs and high-pulse-rate psychophysical thresholds was assessed the correlation was much poorer. The best correlation of 0.75 was shown between a 30-Hz EABR and the 500-Hz psychophysical threshold. This suggests that the low-pulse-rate EABR threshold is a poor predictor of the high-pulse-rate psychophysical threshold used in electrode mapping. Hence, if the EABR is to be a clinical tool for setting the electrodes it should be used cautiously and with the proper correction formula applied.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.