Abstract

Information is lacking about the reliability and agreement of different shear-wave elastographic modes in the peripheral nervous system evaluation. The aim of this observational study was to evaluate reproducibility and agreement of two different shear-wave elastographic modes for measuring the sciatic nerve stiffness in patients affected by osteoarthrosis. Two sets of three measurements were conducted bilaterally on the sciatic nerve of 20 patients with point and 2-D shear-wave elastography by a unique expert sonographer. This consecutive case series study was performed in 1 mo. No significant difference was found comparing the first with the second set of evaluations (p = 0.08 for point shear-wave elastography and 0.3 for 2-D shear-wave elastography). Correlation between the two sets of measurements was good and excellent (0.799 with point shear-wave elastography and 0.877 with 2-D shear-wave elastography). Intra-class coefficient correlation between the two sets of measurements was excellent for both shear-wave elastographic modes (0.869 and 0.938, respectively); no agreement between them was demonstrated (analysis of variance [ANOVA] test: p = 0.014).Despite the lack of agreement owing to the different procedures for measuring, both shear-wave elastographic modes allow reliable stiffness measurements of the sciatic nerve and may be used to evaluate stiffness changes.