Abstract

ObjectivesTo determine the reliability and accuracy of intraoral radiography (IR), cone-beam-computed tomography (CBCT), and dental magnetic resonance imaging (dMRI) in measuring peri‑implant bone defects around single zirconia implants. MethodsTwenty-four zirconia implants were inserted in bovine ribs with various peri‑implant defect sizes and morphologies. True defect extent was measured without implant in CBCT. Defects were measured twice in IR, CBCT, and dMRI with the inserted implant by three experienced readers. Reliability was assessed by ICC, accuracy by the Friedman test, and post-hoc-Tukey's test. ResultsA comparable good to excellent intra- and inter-reader reliability was observed for all modalities (intra-/inter-rater-CC range for IR; CBCT; dMRI: 0.81–0.91/0.79;0.87–0.97/0.96;0.87–0.95/0.94). Accuracy was generally high, with mean errors below 1 mm in all directions. However, measuring defect depth in the mesiodistal direction was significantly more accurate in dMRI (0.65 ± 0.38 mm) compared to IR (2.71 ± 1.91 mm), and CBCT (1.98 ± 1.97 mm), p-values ≤ 0.0001 respectively ≤ 0.01. ConclusionsOsseous defects around zirconia implants can be reliably measured in IR/CBCT/dMRI in the mesiodistal directions. In addition, CBCT and dMRI allow assessment of the buccolingual directions. dMRI provides a comparable accuracy in all directions, except for the mesiodistal defect depth, where it outperforms IR and CBCT.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.