Abstract

In clinical practice, stripping the plastic tubes of intravenous (i.v.) infusion sets with a scissor blade is a commonly used method for re-establishing flow in malfunctioning i.v. sets. The present investigation concludes that this procedure results in release of plastic particles from the luminal wall of the tube. Particles are subsequently flushed into the patient. The average amount of particles released exclusively from the i.v. infusion sets under these circumstances may exceed the standards for acceptable particle content per millilitre in large volume i.v. injectable fluids, according to the British Pharmacopoeia (1980) and the United States Pharmacopoeia (1990).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.