Abstract

The IAEA TRS483 and TRS398 Code of Practices (CoP) were used to calculate relative output factors for small photon beams of 6X, 6XFFF energies shaped by High Definition Multileaf Collimator (HDMLC), jaws and cones mounted on TrueBeam STx medical linear accelerator (Varian Medical Systems), respectively. A comparison between these results were made. The results show a large discrepancy in relative output factor curves found among different collimation systems of the same equivalent field sizes and between the CoPs. Therefore, the specific beam modelling in treatment planning system for each type of the collimation system to be used for small fields maybe required for better computational accuracy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.