Abstract

In recent years, boron neutron capture therapy has received renewed interest as new promising boron compounds and epithermal neutron beams have become available. For establishment of the different dose components in epithermal neutron beams, various techniques are applied, including the use of tissue-equivalent (TE) ionisation chambers for the determination of the intermediate and fast neutron dose. This paper is directed towards the calculation of the relative neutron sensitivity of TE ionisation chambers, kT, for the free-in-air spectrum at the HB11 epithermal beam facility of the High Flux Reactor in Petten. The kT value obtained from the present calculations is 0.87. This value is about 10% lower than data currently applied for similar conditions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.