Abstract

The capability of Timepix3 detectors installed in ATLAS to measure luminosity is evaluated. It is described how noisy pixels are identified and excluded. Two different methods for luminosity determination, i.e. cluster counting and thermal neutron counting are described and compared with each other. The achieved short-term relative precision with both methods is determined by modeling the luminosity curve. It is shown that using cluster counting a short-term relative precision of < 0.5 % can be achieved for 60 s time intervals. For thermal neutrons, a short-term relative precision (for 60 s intervals) of ≈ 2 % was found. Hereby statistics was the limiting factor. The findings are discussed in view of Timepix3 upgrade plans for LHC Run-3.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.