Abstract

The relative energy responses of three indigenously developed optically stimulated luminescence (OSL) phosphors in the disc form were studied in therapeutic photon and electron beams. Calibration in terms of absorbed dose was carried out in the dose range 5-500 cGy in 60 Co gamma rays, high energy X-rays, and electron beams used in radiotherapy. The combined standard uncertainty in the estimation of absorbed dose using these OSL discs (OSLDs) was 3.3%. Dose-response curves of these OSLDs in 60 Co gamma rays, 6 and 10 MV (flat and unflat), 15 MV and 6 and 15 MeV electron beams were found to be linear. Furthermore, these OSLDs exhibited a relative energy-dependent response for both photon and electron beams. The relative energy response correction factor for photon and electron beams were in the range 1.01-1.05 and 1.03-1.06, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.