Abstract
Aided threshold improvement obtained by 12 hearing-impaired children was compared to the acoustic gain of their hearing aids using both the traditional 2-cc coupler and a variable volume coupler designed to approximate real ear volume in children. Results indicated that acoustic gain determined in the 2-cc coupler underestimated aided threshold improvement by approximately 8.7 dB. Use of the variable volume coupler to determine acoustic gain, however, adequately predicted aided improvement at comfort setting. Use of the variable volume coupler in hearing-aid fittings for children is discussed with special emphasis on preventing over-amplification.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.