Abstract

Many ultrasonic parameters, primarily related to attenuation and scatterer size, have been used to characterize the composition of atherosclerotic plaque tissue. In this study, we combine elastographic (axial strain ratio) and ultrasonic tissue characterization parameters, namely the attenuation coefficient and a scattering parameter associated with an “equivalent” scatterer size to delineate between fibrous, calcified, and lipidic plaque tissue. We present results obtained from 44 ex vivo atherosclerotic plaque specimens obtained after carotid endarterectomy on human patients. Our results in the frequency range 2.5∼7.5 MHz indicate that softer plaques (with higher values of the strain ratio) are usually associated with larger equivalent scatterer size estimates (200 ∼500 μm) and lower values of the attenuation coefficient slope (<1 dB/cm/MHz). On the other hand, stiffer plaques (with lower strain ratio values) are associated with smaller equivalent scatterer size estimates (100 ∼200 μm) and higher values of the attenuation coefficient slope (1∼3 dB/cm/MHz). These results indicate that ultrasonic tissue characterization and strain parameters have the potential to differentiate between different plaque types. These parameters can be estimated from radio-frequency data acquired under in vivo conditions and may help the clinician decide on appropriate interventional techniques. (E-mail: tvarghese@wisc.edu)

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call