Abstract

The Quantec Spinal Imaging System (QSIS) is a raster stereography used to measure three-dimensional trunk images. The Q angle, a coronal plane measurement generated by the Quantec Spinal Imaging System (QSIS), was compared with the Cobb angle in assessment of scoliosis curve magnitude. One hundred forty-nine patients with idiopathic scoliosis were evaluated using both the Quantec system and plane radiographs. The Cobb and Q angles demonstrated significant correlation in the thoracic region (r = 0.65, p < 0.05), lumbar region (r = 0.63, p < 0.05), and in the thoracolumbar region (r = 0.70, p < 0.05). The difference between the Q and Cobb angles was small when the Cobb angle was <21 degrees with less than 6 degrees of axial surface rotation, as measured by the QSIS method. For smaller curves with minimal rotation, there is close correlation between the Cobb angle and the Quantec angle.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.