Abstract

To evaluate the micromobility of dental implants under occlusal loading in relation to stability measurements of resonance frequency analysis and insertion torque. The sample comprised of 24 implants inserted in 12 fresh cow ribs. Insertion torque and Osstell implant stability quotient (ISQ) measurements were recorded. An "ad hoc" acrylic premolar was made on a temporary abutment and screwed to each implant, and a force of 100 N was subsequently applied at an angle of 6 degrees. Implant micromotion was measured using a Questar microscope with a resolution of 2 μm and an image analysis program. Data show a statistically significant inverse correlation between the ISQ values and implant micromotion under a load of 100 N (R = 0.86, P < 0.0001). The same relationship is found between insertion torque and implant micromotion, although the relationship is linear up to 34 N·cm and becomes exponential for higher values (R = 0.78, P < 0.0001). A direct correlation is established between insertion torque and ISQ values. There is an inverse relationship between both ISQ and insertion torque values and implant micromotion under a load of 100 N.