Abstract

To explore the relationship between primary implant stability and different parameters related to implant or bone properties. Twenty-four patients received a total of 136 Straumann SLActive implants. Resonance frequency analysis (RFA) was performed at implant placement, and RFA and Periotest (PTV) were scored at loading. Bone density [Hounsfield (HU) scores] and coronal cortical thickness at osteotomy sites were measured from pre-operative computerized tomography scans. Implant length, diameter or the presence of bony dehiscence did not have a significant effect on the mean RFA scores at implant insertion. Significant linear relations were found between RFA or PTV scores and HU values (P<0.05), or cortical bone thickness (P<0.05), both at insertion as well as at loading. RFA and PTV scores can be predicted based on implant and especially bone related factors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.