Abstract

A laser-induced fluorescence set-up profiling physico-chemical parameters of droplets and their changes in an electrospray plume is coupled to an Agilent single quadrupole mass spectrometer equipped with an Agilent Jet Stream electrospray source. This coupling allows investigating the evolution of the pH of droplets as they evaporate in the electrospray plume by measuring emission spectra of a pH-chromic dye. Simultaneous measurement of mass spectra and optical pH profiling permits to study the relation between the charge states of peptide anions in mass spectra and pH changes in the spray plume. In particular, due to changes in droplets evaporation rates, the source parameters (i.e., sheath gas flow and temperature) influence the pH of the droplets and relative intensities of charge states in peptide mass spectra.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.