Abstract
On May 26, 1976, the Food and Drug Administration (FDA) began implementing the Medical Device Amendments to the Federal Food Drug and Cosmetic Act. These amendments give FDA specific authority to regulate "medical devices." Additional authority was provided in the Safe Medical Devices Act of 1990. The pathways to get in vitro diagnostic products to market for commercial distribution include: premarket notification [or 510(k)] and premarket approval (PMA). Reviews of these applications are performed in the Division of Clinical Laboratory Devices, Office of Device Evaluation within the Center for Devices and Radiological Health. Most submissions for invasive glucose devices are 510(k) applications. Through this process the manufacturer demonstrates that its device is substantially equivalent to another device that is already legally marketed, referred to as the predicate device. To date, noninvasive and minimally invasive glucose testings have been approved through the PMA process and will be discussed in a separate article.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.