Abstract

Most Mueller matrix imaging polarimeters implement sequential acquisition of at least 16 raw images of the same object with different incident and detected light polarizations. When this technique is implemented in vivo, the unavoidable motions of the subject may shift and distort the raw images to an extent such that the final Mueller images cannot be extracted. We describe a registration algorithm which solves this problem for the typical conditions of in vivo imaging, e.g. with spatially inhomogeneous medium to strong depolarization. The algorithm, based on the so called "optical flow," is validated experimentally by comparing the Mueller images of a pig skin sample taken in static and in dynamic conditions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.