Abstract

Tomographic registration, a method that makes possible accurate patient registration directly from projection data, consists of three processing steps: (i) manual coarse positioning, (ii) tomographic projection set acquisition, and (iii) computer mediated refined positioning. In the coarse positioning stage, the degree of patient alignment is comparable with that achieved with the standard radiotherapy set-up. However, the accuracy requirements are somewhat more relaxed in that meticulous alignment of the patient using external laser indicators is not necessary. Instead, tomographic projection sets are compared with planning CTs in order to achieve improved patient set-up. The projection sets are cross-correlated to obtain the best-fit translation and rotation offsets. The algorithm has been tested on synthetic data with the incorporation of varying amounts of Gaussian pseudo-random noise. These tests demonstrate the algorithm's stability and also confirm that alignment can be achieved with an accuracy of less than one projection pixel.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.