Abstract
Ochratoxin A (OTA) can contaminate foodstuffs in the ppb to ppm range and once formed, it is difficult to remove. Because of its toxicity and potential risks to human health, the need exists for rapid, efficient detection methods that comply with legal maximum residual limits. In this work we have synthesized an OTA conjugate functionalized with a water-soluble peptide for covalent immobilization on a glass biochip by means of contact spotting. The chip was used for OTA determination with an indirect competitive immunoassay format with flow-through reagent addition and chemiluminescence detection, carried out with the stand-alone automated Munich Chip Reader 3 (MCR 3) platform. A buffer model and real green coffee extracts were used for this purpose. At the present, covalent conjugate immobilization allowed for at least 20 assay-regeneration cycles of the biochip surface. The total analysis time for a single sample, including measurement and surface regeneration, was 12 min and the LOQ of OTA in green coffee extract was 0.3 μg L −1 which corresponds to 7 μg kg −1.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.