Abstract
We examine the effects of aberrations induced by a refractive index mismatch on the signal level and resolution of single-photon (1-p) and two-photon (2-p), conventional and confocal scanning microscopes. In particular, we consider the aberrations introduced by an interface between oil/glass and water. Resolution is defined in terms of enclosed fluorescence, rather than full-width half-maximum, revealing more useful information for heavily aberrated point spread functions (PSFs). It is shown that, at large focusing depths, the resolution of 2-p conventional and 1-p confocal microscopes are almost identical. The benefits of aberration correction are examined by removing Zernike aberration modes. With aberration correction, the best resolution is found for 1-p confocal and 2-p confocal modes. An approximation based upon geometrical optics is also introduced which shows that the axial resolution of heavily aberrated PSFs is roughly proportional to focusing depth.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.